Health product recall

Alphenix INFX-8000C

Last updated

Summary

Product
Alphenix INFX-8000C
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Alphenix INFX-8000C

PFC2237004

INFX-8000C/PF

Issue

It has been found that the c-arm and table may not be able to operate because the soldering on the tableside console printed wiring board was not done properly.

Recall start date: May 8, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Canon Medical Systems Corporation

1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73688

Get notified