Health product recall

ALLURA XPER FD20 (2014-12-09)

Starting date:
December 9, 2014
Posting date:
April 13, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52853

Recalled Products   

ALLURA XPER FD20 SYSTEM - MAIN UNIT

Reason

Philips has become aware that a mechanical noise/tick might occur when moving the XY-carrier over the X-rail coupling. This problem can occur when an "old" version of coupling block with M5 bolts is installed in the system.

Affected products

ALLURA XPER FD20 SYSTEM - MAIN UNIT

Lot or serial number

All lots.

Model or catalog number

ALLURA XPER FD20

 