Allura Xper Fd10 & Fd20 products
Last updated
Summary
Product
Allura Xper Fd10 & Fd20 products
Issue
Medical devices - Mechanical defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products | Lot or serial number | Model or catalog number |
---|---|---|
Allura Xper Fd10/10 (Product Of Alluraclarity Family) | Not applicable | 722027 |
Allura Xper Fd10 (Product Of Alluraclarity Family) | Not applicable | 722026 |
Allura Xper Fd10 | Not applicable | 001443 |
Allura Xper Fd20 (Product Of Alluraclarity Family) | Not applicable | 722028 |
Issue
Philips has discovered a problem with the flat detector cooling system. Due to a leak in the flat detector cooling system, cooling liquid may leak outside the drip tray of the laird chiller. If this occurs, the liquid may drip into electrical components in the R-cabinet located in the technical room which could cause damage to the system.
Recall start date: Nov 22, 2021
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Medical Systems Nederland B.V. |
Veenpluis 6, Best Netherlands, 5684 PC |
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63712
Get notified
Receive notifications for new and updated recalls and alerts by category.