Allura Xper Fd10 & Fd20 products
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
Allura Xper Fd10/10 (Product Of Alluraclarity Family) |
Not applicable |
722027 |
Allura Xper Fd10 (Product Of Alluraclarity Family) |
Not applicable |
722026 |
Allura Xper Fd10 |
Not applicable |
001443 |
Allura Xper Fd20 (Product Of Alluraclarity Family) |
Not applicable |
722028 |
Issue
Philips has discovered a problem with the flat detector cooling system. Due to a leak in the flat detector cooling system, cooling liquid may leak outside the drip tray of the laird chiller. If this occurs, the liquid may drip into electrical components in the R-cabinet located in the technical room which could cause damage to the system.
Recall start date: Nov 22, 2021