Health product recall

ALLURA XPER FD (2015-06-22)

Starting date:
June 22, 2015
Posting date:
July 17, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54236

Affected Products

  1. Allura XPER FD10/10
  2. Allura XPER FD20/10
  3. Allura XPER FD20/20
  4. Allura XPER FD10
  5. Allura XPER FD20

Reason

Philips had received a call with respect to an issue with a failing Monitor Ceiling Suspension (MCS) actuator of an Allura Xper system. The actuator assembly had become detached and the monitor carriage with a FlexVision monitor dropped to its lowest position. The monitor ceiling suspension is designed to allow flexible positioning near the patient table when in use, and away from the patient when not in use. (Park position).

Affected products

A. Allura XPER FD10/10

Lot or serial number

>10, contact mfg.

Model or catalog number

ALLURA XPER FD10/10

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



B.  Allura XPER FD20/10

Lot or serial number

>10, contact mfg.

Model or catalog number

Allura XPER FD20/10

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



C.  Allura XPER FD20/20

Lot or serial number

>10, contact mfg.

Model or catalog number

Allura XPER FD20/20

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



D.  Allura XPER FD10

Lot or serial number

>10, contact mfg.

Model or catalog number

Allura XPER FD10

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



E.  Allura XPER FD20

 

Lot or serial number

>10, contact mfg.

Model or catalog number

Allura XPER FD20

 

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands