Health product recall

Alissa Interpret

Last updated

Summary

Product
Alissa Interpret
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Alissa Interpret v5.3.7 v5.4.3 K5849AA

Issue

1. The ACMG (American College of Medical Genetics and Genomics) classification scheme implemented in Alissa Interpret does not result in variant of uncertain significance (VOUS) classification when benign and pathogenic variant are contradictory.
2. The ACMG classification scheme does not result in the correct classification when the criteria strength is updated.

Recall Start Date: December 21, 2023
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Agilent Technologies Belgium

Technologielaan 3, Leuven, Belgium,

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74986

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe