Aliquot Plates (2014-04-28
- Starting date:
- April 28, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39497
Recalled Products
Aliquot Plates
Reason
Siemens Healthcare Diagnostics investigated several "Aliquot Plate Load Failed" complaints. The green lid stock is visible separating form the white base of the aliquot plates causing the plate to not properly load onto the dimension Vista System. If a defective aliquot plate is placed in one of the three bays of the instrument aliquot plate holders for loading, an error "Aliquot Plate Load Failed” is generated. The user needs to physically remove the plate from the aliquot plate holder and take the necessary steps to clear the error. Unaffected aliquot plates load without an error.
Affected products
Aliquot Plates
Lot or serial number
All lots
Model or catalog number
10445210
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES