Health product recall

Aliquot Plates (2014-04-28

Starting date:
April 28, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39497

Recalled Products

Aliquot Plates

Reason

Siemens Healthcare Diagnostics investigated several "Aliquot Plate Load Failed" complaints.  The green lid stock is visible separating form the white base of the aliquot plates causing the plate to not properly load onto the dimension Vista System. If a defective aliquot plate is placed in one of the three bays of the instrument aliquot plate holders for loading, an error "Aliquot Plate Load Failed” is generated.  The user needs to physically remove the plate from the aliquot plate holder and take the necessary steps to clear the error. Unaffected aliquot plates load without an error.

Affected products

Aliquot Plates 

Lot or serial number

All lots 

Model or catalog number

10445210

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES