This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Alinity I Total PSA Reagent Kit

Last updated

Summary

Product
Alinity I Total PSA Reagent Kit
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Alinity I Total PSA Reagent Kit

73162FZ00
71210FZ00
71210FZ01
71343FZ00
73298FZ00
73155FZ00
71213FZ00
73155FZ01
71346FZ00
73728FZ00

07P9220
07P9230

Issue

An elevated number of complaints have been reported involving both out-of-range third-party quality control (qc) results and a positive bias in patient sample results when using specific product lots. Internal testing with patient samples, including retained lot analysis and trending of patient result medians, confirmed that some lots may exhibit a greater than 10% positive bias.

Recall start date: May 22, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Immunology
Companies

Abbott Ireland, Diagnostics Division

Finisklin Business Park, Sligo, Ireland

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77561

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe