Alinity hq Analyzer, Alinity hs Slide Maker Stainer Module (2019-03-04)
- Starting date:
- March 4, 2019
- Posting date:
- June 14, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70217
Last updated: 2019-06-14
Affected Products
A. Alinity hq Analyzer
B. Alinity hs Slide Maker Stainer Module
Reason
Abbott has identified the safety interlock covering the septum piercing probes may not deploy when the bulk solution 1 Liter bottle is removed on the Alinity h-series: Alinity hq Analyzer & the Alinity hs Slide Maker Stainer Module.
Affected products
A. Alinity hq Analyzer
Lot or serial number
All serial numbers.
Model or catalog number
09P68-01
Companies
- Manufacturer
-
Abbott GMBH & Co. K.G.
MAX PLANCK RING 2
WIESBADEN
65205
GERMANY
B. Alinity hs Slide Maker Stainer Module
Lot or serial number
All serial numbers.
Model or catalog number
09P69-01
Companies
- Manufacturer
-
Abbott GMBH & Co. K.G.
MAX PLANCK RING 2
WIESBADEN
65205
GERMANY