Alinity HQ Analyzer
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Alinity HQ Analyzer |
More than 10 numbers, contact manufacturer. |
09P68-01 |
Issue
Abbott has identified the following issues when using Alinity HQ Analyzer software version 5.6 and below which will be corrected in software version 5.9.
Issue 1: When processing a CBC+diff sample that immediately follows a CBC+diff+retic sample, diluent may backflow into the RBC dilution cup. This may result in falsely low red blood cell (RBC) results and consequently falsely high mean cell hemoglobin concentration (MCHC), falsely high mean cell hemoglobin (MCH), and falsely low hematocrit (HCT).
Issue 2: An overestimation of the basophil (baso) count may occur on some samples when cell events are incorrectly counted as baso during sample analysis. this may result in falsely increased baso counts and %b (%baso).
Recall start date: June 21, 2024