Alinity i Free T3 Reagent Kit and Alinity I Total T3 Reagent Kit (2018-10-16)
- Starting date:
- October 16, 2018
- Posting date:
- November 2, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68174
Affected products
- Alinity i Free T3 Reagent Kit
- Alinity i Total T3 Reagent Kit
Reason
During Non-Conformance investigation a Reagent Cross Contamination (RCC) has been observed for the 6 and 2-
point ARCHITECT Free T3 and Total T3 assays on the i1000, when tested alongside assays containing poly-l-lysine (pll) e.g. ARCHITECT TSH. RCC has also been observed for the 2 and 6-point ARCHITECT free T3 and total T3 assays when tested on the i2000/i2000SR system and the Alinity i System.
Affected products
A. Alinity i Free T3 Reagent Kit
Lot or serial number
All lots.
Model or catalog number
- 07P6920
- 07P6930
Companies
- Manufacturer
-
Abbott Ireland Diagnostic Division
Lisnamuck, Longford
IRELAND
B. Alinity i Total T3 Reagent Kit
Lot or serial number
All lots.
Model or catalog number
- 07P9420
- 07P9430
Companies
- Manufacturer
-
Abbott Ireland Diagnostic Division
Lisnamuck, Longford
IRELAND