Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71563


Last updated:

Affected Products

Alinity c Carbon Dioxide Reagent Kit

Reason

Abbott internal studies have determined that atmospheric carbon dioxide (CO2) can be absorbed into the Alinity c carbon dioxide reagent, resulting in the potential for incorrect results.  This testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results.

Affected products

Alinity c Carbon Dioxide Reagent Kit

Lot or serial number
  • 54449UQ03
  • 54454UQ03
Model or catalog number
  • 07P7220
  • 07P7230