Health product recall

Alcon or Endure LuxOR E71 Ophthalmic Microscope and Q-Vue Ophthalmic Microscope (2018-07-18)

Starting date:
July 18, 2018
Posting date:
August 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67344

Affected products

  1. Alcon or Endure LuxOR E71 Ophthalmic Microscope
  2. Alcon or Endure LuxOR E71 Q-Vue Ophthalmic Microscope

Reason

Through our monitoring process, Alcon has observed there is a potential for the optical head to detach from the stand on the Alcon or Endure LuxOR E71 Ophthalmic Microscope and Alcon or Endure LuxOR E71 Q-Vue Ophthalmic Microscope due to the lack of thread adhesive. This adhesive is designed to prevent unthreading of the mounting post.

Affected products

A. Alcon or Endure LuxOR E71 Ophthalmic Microscope

Lot or serial number
  • s/n: 1403918102X
  • s/n: 1502158505X
Model or catalog number

8065752322

Companies
Manufacturer

Alcon Laboratories Inc.

6201 South Fwy

Fort Worth

76134

Texas

UNITED STATES



B. Alcon or Endure LuxOR E71 Q-Vue Ophthalmic Microscope

Lot or serial number
  • s/n: 1402314404X
  • s/n: 1402314405X
  • s/n: 42119:001
  • s/n: 42994:001
  • s/n: 44040:001
Model or catalog number

8065752240

Companies
Manufacturer

Alcon Laboratories Inc.

6201 South Fwy

Fort Worth

76134

Texas

UNITED STATES