Alaris™ Pump Module & BD Alaris™ Pump Module
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Alaris™ Pump Module & BD Alaris™ Pump Module |
More than 10 numbers, contact manufacturer. |
8100 |
Issue
BD has become aware via customer complaints of customer/third-party continued usage of previously recalled bezel kit assemblies to service Alaris™ and BD Alaris™pump modules. These affected bezel kit assemblies were manufactured with fr-110 resin between April 2011 and June 2017 and should not be used in servicing as these may present with separated and/or broken bezel bosses that can lead to free flow, over infusion, under infusion or interruption of infusion.
Recall start date: July 16, 2025
Additional information
Details
| Carefusion 303, Inc. |
| 10020 Pacific Mesa Blvd., San Diego, California, United States, 92121 |
Get notified