Alaris System PC Unit Models 8000 and 8015
- Starting date:
- November 1, 2016
- Posting date:
- November 18, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61158
Reason
If the PC unit is running on battery power, a low battery alarm and/or very low battery alarm should activate when 30 minutes and 5 minutes of estimated battery runtime remains. In some cases, these low battery alarms will not be generated before the PC unit displays a battery discharged alarm and all infusion channels are stopped.
Affected products
Alaris System PC Unit Models 8000 and 8015
Lot or serial number
Plus de 10 numéros, contactez le fabricant.
Model or catalog number
- 8000
- 8015
Companies
- Manufacturer
-
Carefusion 303, Inc.
10020 Pacific Mesa Blvd.
San Diego
92121-2733
California
UNITED STATES