Alaris System PC Unit Model 8015
- Starting date:
- June 12, 2017
- Posting date:
- July 6, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-63852
Reason
System error 255-16-275 can occur when a user selects two functions at the same time/rapid succession or not following typical workflows. This results in a synchronization issue between the pc unit and the modules.
Affected products
Alaris System PC Unit Model 8015
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
8015
Companies
- Manufacturer
-
Carefusion 303, Inc.
10020 Pacific Mesa Blvd.
San Diego
92121-2733
UNITED STATES