This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Alaris System - Alaris Server Mobile (2016-03-03)
- Starting date:
- March 3, 2016
- Posting date:
- April 22, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-57998
Affected products
A. Alaris System - Alaris Server Mobile
Reason
CareFusion has identified a potential risk with the Alaris System model 8015 manufactured between September 11, 2012 and March 29, 2013. The units manufactured during this time frame may display a system error code 133.6080 due to a failure with the super capacitor (C245) at power up on the Alaris System logic boards (part numbers TC 10007253 and TC 100006939).
Affected products
A. Alaris System - Alaris Server Mobile
Lot or serial number
8015
Model or catalog number
SERIAL #S 12706344
SERIAL #S 12992376
SERIAL #S 13691087
SERIAL #S 137XXXXX > 100
SERIAL #S 138XXXX >100
Companies
- Manufacturer
-
Carefusion 303, Inc.
10020 Pacific Mesa Blvd.
San Diego
92121-2733
California
UNITED STATES