This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Alaris System - Alaris Server Mobile (2016-03-03)

Starting date:
March 3, 2016
Posting date:
April 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-57998

Affected products

A. Alaris System - Alaris Server Mobile

Reason

CareFusion has identified a potential risk with the Alaris System model 8015 manufactured between September 11, 2012 and March 29, 2013. The units manufactured during this time frame may display a system error code 133.6080 due to a failure with the super capacitor (C245) at power up on the Alaris System logic boards (part numbers TC 10007253 and TC 100006939).

Affected products

A. Alaris System - Alaris Server Mobile

Lot or serial number

8015

Model or catalog number

SERIAL #S 12706344

SERIAL #S 12992376

SERIAL #S 13691087

SERIAL #S 137XXXXX > 100

SERIAL #S 138XXXX >100

Companies
Manufacturer

Carefusion 303, Inc.

10020 Pacific Mesa Blvd.

San Diego

92121-2733

California