This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Alaris Syringe Model 8110 (2015-07-09)
- Starting date:
- July 9, 2015
- Posting date:
- August 24, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54776
Affected Products
A. Alaris Syringe Model 8110
Reason
After error code 351.6740 and channel error is cleared on the PCU, the syringe pump is unresponsive to key presses until the next power cycle, or the module is detached and re-attached.
Affected products
A. Alaris Syringe Model 8110
Lot or serial number
contact manufacturer
Model or catalog number
8110
Companies
- Manufacturer
-
Carefusion 303, Inc.
10020 Pacific Mesa Blvd.
San Diego
92121-2733
California
UNITED STATES