This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Alaris Syringe Model 8110 (2015-07-09)

Starting date:
July 9, 2015
Posting date:
August 24, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54776

Affected Products

A. Alaris Syringe Model 8110

Reason

After error code 351.6740 and channel error is cleared on the PCU, the syringe pump is unresponsive to key presses until the next power cycle, or the module is detached and re-attached.

Affected products

A. Alaris Syringe Model 8110

Lot or serial number

contact manufacturer

Model or catalog number

8110

Companies
Manufacturer

Carefusion 303, Inc.

10020 Pacific Mesa Blvd.

San Diego

92121-2733

California

UNITED STATES