This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

AISYS CS2 ANESTHESIA MACHINE MAIN UNIT (2016-03-10)

Starting date:
March 10, 2016
Posting date:
April 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-57770

Affected Products

AISYS CS2 ANESTHESIA MACHINE MAIN UNIT

Reason

The vacuum suction tubing could be kinked within the machine resulting in suction less than the 20 LPM required by ISO 10079-3. If this issue is left unresolved, it could result in inadequate suction performance to clear the airway and inadequate ventilation of the patient. There have been no injuries reported as a result of this issue.

Affected products

AISYS CS2 ANESTHESIA MACHINE MAIN UNIT

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

1011-9050-000

Companies
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES