Aisys Cs2
Brand(s)
Last updated
Summary
Product
Aisys Cs2
Issue
Medical devices - Performance
What to do
Continue using anesthesia system if no cracks. If crack observed, contact manufacturer.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Aisys Cs2 |
More than 10 numbers, contact manufacturer. |
1011-9050-000 |
Issue
The affected anesthesia systems base can have a crack in a specific location in the rear of the system. If excessive load is applied to the device, a cracked base could potentially fracture, resulting in a tip or overbalance of the anesthesia system, which can result in potential injury.
Recall start date: June 3, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Datex-Ohmeda, Inc.
3030 Ohmeda Drive, Madison, Wisconsin, United States, 53707-7550
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64266
Get notified