This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Affinity Four Birthing Beds P3700B (2014-06-20)

Starting date:
June 20, 2014
Posting date:
July 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40281

Recalled Products

Affinity Four Birthing Beds P3700B

Reason

Hill-Rom has received thirteen reports over the past 8 year from facilities in the USA alleging that the Affinity Four lift-off foot section has fallen resulting in three minor injuries. Analysis suggests that the latch mechanism may have been bent due to user error during handling of the foot section, such as dropping it to the floor.

Affected products

Affinity Four Birthing Beds P3700B 

Lot or serial number

Not applicable

Model or catalog number

P3700B

Companies
Manufacturer

Hill-Rom Company, Inc.

1069 State Route 46 East

Batesville

47006-9167

Indiana

UNITED STATES