This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

AFFINITY FOUR BIRTHING BEDS (2016-07-11)

Starting date:
July 11, 2016
Posting date:
July 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-59392

Affected products

A. AFFINITY FOUR BIRTHING BEDS

Reason

Potentially defective head drives (a plastic carry arm within the head drive is defective). If the failure mode does occur, it could prevent the CPR quick release mechanism from operating properly

Affected products

A. AFFINITY FOUR BIRTHING BEDS

Lot or serial number
  • "R039AA7278
  • R039AA7279
  • R039AA7281
  • R039AA7284
  • R046AA7345"
Model or catalog number
  • P3700E
Companies
Manufacturer

Hillrom Corporation

1069 State Route 46 East

Batesville

47006

UNITED STATES