Health product recall

ADVIA Chemistry System - Urea Nitrogen (UN) Assay (2018-04-05)

Starting date:
April 5, 2018
Posting date:
May 1, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66632

Affected products

ADVIA Chemistry System - Urea Nitrogen (UN) Assay

Reason

Siemens Healthcare Diagnostics has confirmed that the Urea Nitrogen Reagent Kit Lots 408143 and 418290 for use on ADVIA 1800, 2400 and XPT Chemistry Systems demonstrate an increased incidence of calibration failure. Results cannot be generated by the system when calibration fails. If calibration passes, there is a potential for falsely elevated patient and quality control results. Internal testing has also shown that results have the potential to be falsely increased by approximately 10 mg/dl (3.57 mmol/l) across the analytical range. Quality controls may not always detect the elevated results.

Affected products

ADVIA Chemistry System - Urea Nitrogen (UN) Assay

Lot or serial number

408143

Model or catalog number

03040257

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES