Health product recall

ADVIA CHEMISTRY OPIATE ASSAYS

Starting date:
August 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60074

Affected Products

  1. ADVIA CHEMISTRY OPIATE (OP_2) ASSAY
  2. ADVIA 1200 CHEMISTRY SYSTEM- OPIATE (OP_2) ASSAY2

Reason

Siemens investigation confirmed that the introduction of a new SYVA polyclonal antibody lot (G1) for use in the dimension urine opiates screen has resulted in an alteration of specificity for Levallorphan, Levorphanol, Nalorphine and Oxycodone at the 300 and 2000 NG/ml cutoff concentration. As a result in the change in sensitivity to the cross reactant, the claims for these compounds are no longer accurate for either the 300 or 2000 NG/ml cutoff as stated within the instructions for use (IFU). Further investigation confirmed that SYVA, Beckman Coulter and Advia Chemistry Opiate reagents are impacted as well due to a common raw material used in the assays.

Affected products

A. ADVIA CHEMISTRY OPIATE (OP_2) ASSAY

Lot or serial number
  • 290
  • 310
  • 320
  • 330
  • 340
Model or catalog number
  • 10378763
  • Companies
    Manufacturer

    Siemens Healthcare Diagnostics Inc.

    511 Benedict Ave

    Tarrytown

    10591

    New York

    UNITED STATES



B. ADVIA 1200 CHEMISTRY SYSTEM- OPIATE (OP_2) ASSAY2

Lot or serial number
  • 290
  • 310
  • 320
  • 330
  • 340
Model or catalog number
  • 10378763
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES