ADVIA Centaur XPT Instrument (2016-06-21)
- Starting date:
- June 21, 2016
- Posting date:
- June 30, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-59124
Affected products
A. ADVIA Centaur XPT Instrument
Reason
Siemens Healthcare Diagnostics is issuing a customer notification for ADVIA Centaur XPT Waste Reservoir Assembly (SMN 10711091). Siemens has determined that certain reservoirs manufactured with walls that are below the nominal specification, potentially cause cracking & leaking of seam bottom and leaking at the interface of the seam and reservoir. If leak occurs foaming may occur in the glass jar in the bulk fluid drawer and cause the instrument to stop with low vacuum error #016001324.
Affected products
A. ADVIA Centaur XPT Instrument
Lot or serial number
IRL06591516
IRL06641516
IRL06661516
IRL06741517
IRL06971520
IRL08721541
IRL08751541
IRL08781542
IRL09631550
IRL09651550
Model or catalog number
10711433
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES