This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advia Centaur System – Instruments (Monthly and Daily Cleaning Procedure)

Starting date:
August 16, 2012
Posting date:
September 3, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-23851

Recalled products

  1. Advia Centaur System – Instrument
  2. Advia Centaur XP System – Instrument
  3. Advia Centaur XP System – Instrument
  4. Advia Centaur XP System – Instrument
  5. Advia Centaur System – Instrument
  6. Advia Centaur System – Instrument

Reason

The Monthly Cleaning Procedure (MCP) maintenance, under certain circumstances, may result in residual cleaning solution (Bleach) in the Reagent Probe 1 (RP1) fluidics and the water reservoir causing false results. Tests affected are Total Human Chorionic Gonadotropin (THCG), Progesterone and Prostate Specific Antigen (PSA). The Daily Cleaning Procedure (DCP) under certain circumstances may result in residual cleaning solution (Bleach) in the water line causing false results that use water for test processing. Test affected is Toxoplasma M (Toxo M).

Affected products

A. Advia Centaur System – Instrument

Lot or serial number

N/A

Model or catalog number

04911855

078-A001-14

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES



B. Advia Centaur XP System – Instrument

Lot or serial number

N/A

Model or catalog number

078-A010-07

09066916

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES



C. Advia Centaur XP System – Instrument

Lot or serial number

N/A

Model or catalog number

078-A010-07

09066916

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES



D. Advia Centaur XP System – Instrument

Lot or serial number

N/A

Model or catalog number

078-A010-07

09066916

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES



E. Advia Centaur System – Instrument

Lot or serial number

N/A

Model or catalog number

04911855

078-A001-14

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES



F. Advia Centaur System – Instrument

Lot or serial number

N/A

Model or catalog number

04911855

078-A001-14

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES