This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advia Centaur System-Calibrators

Starting date:
January 12, 2017
Posting date:
February 1, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62060

Reason

Siemens Healthcare Diagnostics has confirmed a negative bias for Advia Centaur FT4 when used with Calibrator A Kit lots ending in 90 on the Advia Centaur, Advia Centaur XP and Advia Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator relative light units (RLU) %cvs when using Calibrator A Kit lots ending in 90 with the FT4 assay.

The performance of the FT4 assay when used with Calibrator A Kit lots ending in 90 on the Advia Centaur CP System is not affected.

Customers may continue to use Calibrator A Kit lots ending in 90 for the Advia Centaur Systems FT3, T3, T4 and TUP assays.

Internal investigations were performed using available FT4 lots. The investigation confirmed an overall negative bias when comparing calibrator a kit lots ending in 90 to previously released calibrator a kit lots ending in 89.

In some instances the negative bias may cause euthyroid patient samples to result low and outside the reference interval listed in the instructions for use (IFU).
Quality control material and master curve material may result low and outside acceptable ranges.

Affected products

Advia Centaur System-Calibrators

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

04800646
04800735

Companies

Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES