This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Advia Centaur System-Calibrators
- Starting date:
- January 12, 2017
- Posting date:
- February 1, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62060
Reason
Siemens Healthcare Diagnostics has confirmed a negative bias for Advia Centaur FT4 when used with Calibrator A Kit lots ending in 90 on the Advia Centaur, Advia Centaur XP and Advia Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator relative light units (RLU) %cvs when using Calibrator A Kit lots ending in 90 with the FT4 assay.
The performance of the FT4 assay when used with Calibrator A Kit lots ending in 90 on the Advia Centaur CP System is not affected.
Customers may continue to use Calibrator A Kit lots ending in 90 for the Advia Centaur Systems FT3, T3, T4 and TUP assays.
Internal investigations were performed using available FT4 lots. The investigation confirmed an overall negative bias when comparing calibrator a kit lots ending in 90 to previously released calibrator a kit lots ending in 89.
In some instances the negative bias may cause euthyroid patient samples to result low and outside the reference interval listed in the instructions for use (IFU).
Quality control material and master curve material may result low and outside acceptable ranges.
Affected products
Advia Centaur System-Calibrators
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
04800646
04800735
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES