ADVIA Centaur NT-proBNP (PBNP) (2019-12-24)
- Starting date:
- December 24, 2019
- Posting date:
- January 24, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72173
Last updated: 2020-01-24
Affected Products
ADVIA Centaur NT-proBNP (PBNP)
Reason
Siemens Healthcare Diagnostics Inc. confirmed an average method comparison bias of up to 20% with ADVIA Centaur NT-proBNP (PBNP) Assay kit. Siemens’ investigation verified that the ADVIA Centaur PBNP assay continues to meet the reference intervals listed in the ADVIA Centaur PBNP Instructions for Use (IFU) and concordance (agreement) to the predicate method is greater than or equal to 95%. Siemens is currently investigating root cause of the method comparison bias observed and will send a follow-up communication when further information is available.
Affected products
ADVIA Centaur NT-proBNP
Lot or serial number
26878023
Model or catalog number
10699449
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES