Health product recall

ADVIA Centaur and ADVIA Centaur CP System - Creatine Kinase-MB (CKMB) Assay (2018-07-27)

Starting date:
July 27, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67398

Affected products

  1. ADVIA Centaur System-Creatine Kinase-MB (CKMB) Assay
  2. ADVIA Centaur CP System-Creatine Kinase-MB (CKMB) Assay

Reason

Siemens Healthcare Diagnostics has identified a bias with ADVIA Centaur CKMB Calibrator Kit Lots ending in 64 (ck64) when compared to internal standards and previously released ADVIA Centaur CKMB Calibrator Kit Lots ending in 63 (ck63) on the ADVIA Centaur/XP/XPT and ADVIA Centaur CP systems.

Alignment to the internal standardization is restored with the release of ADVIA Centaur CKMB Calibrator Kit Lots ending in 68 (ck68). Customers may observe a shift in Quality Control (QC) material, Master Curve Material (MCM) and patient results when transitioning from ADVIA Centaur CKMB Calibrator Kit Lots ending in 64. Siemens has also identified the potential for out of range ADVIA Centaur CKMB Master Curve Material results when using the lot 35061on the ADVIA Centaur/XP/XPT and ADVIA Centaur CP systems.

Affected products

A. ADVIA Centaur System-Creatine Kinase-MB (CKMB) Assay

Lot or serial number
  • 57834A64
  • 57835A64
  • 72519A64
  • 98115A64

     

Model or catalog number

9318028

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES



B. ADVIA Centaur CP System-Creatine Kinase-MB (CKMB) Assay

Lot or serial number

57834A64

57835A64

72519A64

98115A64

Model or catalog number

9318028

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES