Health product recall

Adult Nasal Cannula

Last updated

Summary

Product
Adult Nasal Cannula
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Adult Nasal Cannula

More than 10 numbers, contact manufacturer.

R-1510
R-1500S

Issue

CHS Ltd. has received customer feedback indicating they had experienced partial blockage of the adult nasal cannula prongs which was impeding the flow of oxygen to the patient. As a result of this situation, CHS conducted an investigation and has been able to confirm this defect. Therefore, CHS is issuing this recall to address these concerns. While there has been no reports of patient injuries to date, CHS has identified this as a type II recall.

Recall Start Date: October 17, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Canadian Hospital Specialties Ltd. Doing Business As Chs

2760 Brighton Road, Oakville, Ontario, Canada, L6H 5T4

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76417

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe