Health product recall

ADULT ANESTHESIA BREATHING CIRCUIT (2021-07-14)

Starting date:
July 14, 2021
Posting date:
July 29, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76097

Last updated:
2021-07-30

Affected Products 

ADULT ANESTHESIA BREATHING CIRCUIT

Reason

Westmed received several complaints pertaining to leaks and holes in the tubing. The issue of leakage was caused due to a sporadic drag along the pulley guide that caused perforations in the tubing.

Affected products

ADULT ANESTHESIA BREATHING CIRCUIT

Lot or serial number

011321T05
012621T37
012621T38
012621T39
021721T15

Model or catalog number

9422

Companies
Manufacturer

Westmed, Inc.

5580 S. NOGALES HIGHWAY

TUCSON

85706

Arizona