This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME (2016-03-08)
- Starting date:
- March 8, 2016
- Posting date:
- March 17, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Hospitals, Healthcare Professionals
- Identification number:
- RA-57574
Affected Products
A. ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME
Reason
Due to an alignment issue with the bezel on the front panel, the plastic cover around the transducer ports prevents full contact of the connector on the transducer to the transducer port on the system. This issue causes the system to lock up.
Affected products
A. ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME
Lot or serial number
400111
Model or catalog number
10433816
Companies
- Manufacturer
-
Siemens Medical Solutions USA
Inc.
685 East Middlefield Road
California