Health product recall

ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME (2016-03-08)

Starting date:
March 8, 2016
Posting date:
March 17, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-57574

Affected Products

A. ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME

Reason

Due to an alignment issue with the bezel on the front panel, the plastic cover around the transducer ports prevents full contact of the connector on the transducer to the transducer port on the system. This issue causes the system to lock up.

Affected products

A. ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME

Lot or serial number

400111

Model or catalog number

10433816