This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ACUSON SC2000 ULTRASOUND - OPERATING SOFTWARE (2016-05-02)

Starting date:
May 2, 2016
Posting date:
May 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-58400

Affected products

  1. ACUSON SC2000 ULTRASOUND - OPERATING SOFTWARE

Reason

While imaging with a transesophageal (TEE) transducer (z6ms, v5ms, or v7m), the operator may lose the ability to control the color region of interest, the pulsed wave or continuous wave Doppler gate or cursor, the M-Mode cursor, the 2D field of view, or the res region of interest with the trackball.

Affected products

A.

ACUSON SC2000 ULTRASOUND - OPERATING SOFTWARE

Lot or serial number
  • 402544
  • 402618
Model or catalog number
  • 4.0
Companies
Manufacturer

Siemens Medical Solutions USA

Inc.

685 East Middlefield Road

California

UNITED STATES