This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Acuson P300 (2014-03-12)

Starting date:
March 12, 2014
Posting date:
April 4, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38687

Recalled products

  1. Acuson P300 

Reason

Using the PA230 tranducer with user-defined preset which includes the 100% setting for output power, could cause the transducer surface temperature to exceed the maximum allowable limit.

Affected products

A. Acuson P300

Lot or serial number
  • 386
  • 497
  • 749
Model or catalog number
  • 10852646
Companies
Manufacturer

Siemens Medical Solutions USA, Inc.

685 East Middlefield Road

Mountain View

94043

California

UNITED STATES