Health product recall

ACTIV.A.C Therapy Unit (2020-06-25)



Starting date:
June 25, 2020
Posting date:
July 10, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73509


Last updated:

Affected Products

ACTIV.A.C Therapy Unit

Reason

KCI has become aware that in certain instances the ACTIV.A.C Therapy Unit may shut down without notification to the user (i.e. no alarm or warning) and the user is not aware that the device has shut down.

Affected products

ACTIV.A.C Therapy Unit

Lot or serial number

Not applicable.

Model or catalog number

340013

Companies
Manufacturer

KCI USA, Inc.

12930 IH 10 WEST

San Antonio

78249

Texas

UNITED STATES