Health product recall

Actim Prom And Actim Partus

Brand(s)
Last updated

Summary

Product
Actim Prom And Actim Partus
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Actim Prom All lots. 30831ETAL 30832ETAL
Actim Partus All lots. 31931ETAL

Issue

Our recent interference testing has shown that the presence of lubricants in the vaginal specimen may interfere with the prom test result, which deviates from what is presented in the instructions for use. The defectiveness was discovered 2023-11-08. This is not a lot specific issue.

Recall Start Date: November 12, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Actim Oy

Klovinpellontie 3, Espoo, Finland, 02180

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74671

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe