Reason

AngioDynamics, Inc., the manufacturer of the ACCU2i PMTA applicator, is conducting a medical device recall on specific lots of these devices. AngioDynamics is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection.

Affected products

ACCU2i PMTA Applicator

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

900-600

900-601

900-602

Companies

Manufacturer

ANGIODYNAMICS, INC.,

No. 1 Falcon Court, Parklands Business Park, Denmead, Hants

Hampshire

P07 6XP

UNITED KINGDOM