This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
ACCU2i PMTA Applicator
- Starting date:
- June 5, 2017
- Posting date:
- July 6, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-63848
Reason
AngioDynamics, Inc., the manufacturer of the ACCU2i PMTA applicator, is conducting a medical device recall on specific lots of these devices. AngioDynamics is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection.
Affected products
ACCU2i PMTA Applicator
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
900-600
900-601
900-602
Companies
- Manufacturer
-
ANGIODYNAMICS, INC.,
No. 1 Falcon Court, Parklands Business Park, Denmead, Hants
Hampshire
P07 6XP
UNITED KINGDOM