Health product recall

Accu-Chek Guide

Last updated

Summary

Product
Accu-Chek Guide
Issue
Medical devices - Labelling and packaging
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalogue number

Accu-Chek Guide

Lot 406670
Lot 406711
SN 93040301566
SN 93040300107
SN 93040304746
SN 93040305146

07398832004

Issue

The issue arises when a blood glucose measurement is taken and the result is displayed in the wrong measuring unit (i.e., mg/dl instead of mmol/l). If it goes unnoticed, the incorrectly displayed blood glucose value may result in inappropriate therapy decisions, including the administration of insulin, and could potentially result in severe consequences. Such possible consequences range from no clinical impact to adverse health events such as hypoglycemia.

Recall start date: May 14, 2024
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies
Roche Diabetes Care Gmbh
Sandhofer Strasse 116, Mannheim, Germany, 68305
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-75885