Access System - Cortisol Assay, UniCel DxI 600 and 800 Access Immunoassay System - Cortisol Assay (2018-09-16)

Starting date:
September 16, 2018
Posting date:
October 19, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67932

Affected products

  1. Access System - Cortisol Assay
  2. UniCel DxI 600 Access Immunoassay System - Cortisol Assay
  3. UniCel DxI 800 Access Immunoassay System - Cortisol Assay

Reason

Beckman Coulter has determined through internal investigation that cross-contamination may have occurred between wells for the affected access cortisol reagent pack lot. Patient results may be negatively impacted (falsely decreased) by up to 16% as a result of this defect.