Access PCT
Last updated
Summary
Product
Access PCT
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Access PCT |
439740 |
C22593 |
Issue
Beckman Coulter has determined that the affected Access PCT reagent lot is experiencing a high rate of calibration failures with a failure code of "max iterate" on Access 2 and UniCel DxI immunoassay systems.
Recall start date: December 30, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Immunotech S.A.S., A Beckman Coulter Company |
130 Ave De Lattre De Tassigny,Bp177, Marseille, France, 13276 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76829
Get notified
Receive notifications for new and updated recalls and alerts by category.