Health product recall

Access PCT

Last updated

Summary

Product
Access PCT
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Access PCT

439740

C22593

Issue

Beckman Coulter has determined that the affected Access PCT reagent lot is experiencing a high rate of calibration failures with a failure code of "max iterate" on Access 2 and UniCel DxI immunoassay systems.

Recall start date: December 30, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Immunotech S.A.S., A Beckman Coulter Company
130 Ave De Lattre De Tassigny,Bp177, Marseille, France, 13276
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76829

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe