Health product recall

Access Intact PTH

Last updated

Summary

Product
Access Intact PTH
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Access Intact PTH 339071 A16972

Issue

Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0.

Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may:
- Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%.
- Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%.
- Produce a zero or near-zero result.

Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment.

Recall Start Date: November 25, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-76621

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe