Access Intact PTH
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Access Intact PTH | 339071 | A16972 |
Issue
Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0.
Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may:
- Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%.
- Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%.
- Produce a zero or near-zero result.
Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment.
Recall Start Date: November 25, 2024
Additional information
Details
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Get notified
Receive notifications for new and updated recalls and alerts by category.