Health product recall

Access hsTnI Reagent (2018-10-15)

Starting date:
October 15, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68382

Affected products

Access hsTnI Reagent

Reason

The Cardiac Troponin T Cross Reactivity concentration of 253 ng/ml (253,000 pg/ml) previously listed in the access HSTNI Instructions For Use (IFU) does not apply to samples with HSTNI concentrations less than 16 pg/ml (ng/l).

Samples containing cardiac troponin t concentrations > 125 ng/ml (125,000 pg/ml) may cross react with the access HSTNI assay:

  • access HSTNI results may change by more than 10% for HSTNI sample concentrations greater than 11.5 pg/ml (ng/l);
  • access HSTNI results may change by more than 2.30 pg/ml (ng/l) for HSTNI sample concentrations less than or equal to 11.5 pg/ml (ng/l).