Access hsTnI (2020-04-08)
- Starting date:
- April 8, 2020
- Posting date:
- May 8, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72997
Last updated: 2020-05-08
Affected Products
Access hsTnI
Reason
Beckman Coulter has determined through customer feedback and an internal investigation that testing a sample with cardiac troponin I (CTNI) concentrations that are > 270,000 pg/ml (ng/L) may cause carryover into the hsTnI reagent pack.
Affected products
Access hsTnI
Lot or serial number
All lots.
Model or catalog number
B52699
Companies
- Manufacturer
-
Immunotech SAS
130 AVENUE DE LATTRE DE TASSIGNY
MARSEILLE CEDEX 09,
13276
FRANCE