Health product recall

Access hsTnI (2020-04-08)



Starting date:
April 8, 2020
Posting date:
May 8, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72997


Last updated:

Affected Products

Access hsTnI

Reason

Beckman Coulter has determined through customer feedback and an internal investigation that testing a sample with cardiac troponin I (CTNI) concentrations that are > 270,000 pg/ml (ng/L) may cause carryover into the hsTnI reagent pack.

Affected products

Access hsTnI

Lot or serial number

All lots.

Model or catalog number

B52699

Companies
Manufacturer

Immunotech SAS

130 AVENUE DE LATTRE DE TASSIGNY

MARSEILLE CEDEX 09,

13276

FRANCE