Health product recall

Access 2 Immunoassay Analyzer

Last updated

Summary

Product
Access 2 Immunoassay Analyzer
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Access 2 Immunoassay Analyzer

More than 10 numbers, contact manufacturer.

81600N

Issue

Beckman Coulter has determined that the Access 2 wash arm lead screw nut band is susceptible to breakage. An Access 2 Immunoassay system wash arm may not move as intended if the nut band is broken. A broken nut band may increase the risk of erroneous results.

Recall start date: January 18, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72633

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe