Health product recall

ABL90 FLEX - Analyzer (2019-06-28)

Starting date:
June 28, 2019
Posting date:
July 19, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70463



Last updated: 2019-07-19

Affected Products

ABL90 FLEX - Analyzer

Reason

On May 14th 2019, Microsoft published information about a vulnerability in Windows XP embedded. The vulnerability is registered as CVE-2019-0708, also called Bluekeep. An attacker could shut down or reboot the device resulting in delayed medical treatment.

Affected products

ABL90 FLEX - Analyzer

Lot or serial number

Softw. version:RXPE v1.3xxxxxx

Model or catalog number

393-090

Companies
Manufacturer

Radiometer Medical Aps

Akandevej 21

Bronshoj

2700

DENMARK