570193 Platinum (DRF)
- Starting date:
- November 27, 2016
- Posting date:
- December 13, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61408
Reason
With the table in vertical position and a displacement request of the table or longitudinal panel to go down with the joystick or a pre-programmed position, the table goes down and continues its movement until the floor.
Affected products
570193 Platinum (DRF)
Lot or serial number
C116F0007
Model or catalog number
6F