3T MRI Systems
Last updated
Summary
Product
3T MRI Systems
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
MR 7700 | Not applicable. | 782120 |
Ingenia Elition X | Not applicable. | 782119 782107 782136 781358 |
Ingenia Elition S | Not applicable. | 782106 782137 781357 |
Issue
Philips has identified an issue where a specific component failure in the gradient coil of where the affected 3T MR Systems identified above may act as a heat source, with a potential to produce smoke and/or fire.
If the components failure occurs, the user may observe the following:
‐ Severe image quality issues (image artifacts caused by electrical disturbances)
‐ Smoke and/or fire
Recall start date: November 22, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands, 5684 PC
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-72329
Get notified