3m™ Steri-Drape™ Surgical Drapes
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| 3m™ Steri-Drape™ Surgical Drapes | More than 10 numbers, contact manufacturer. | 1020NS 1030NS 1033 1060NS 1071 1021NS 1030 1000NS 1021 1060 1061 1010 1010NS 1020 1061NS 1000 |
Issue
During a recent investigation, 3m confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Although the acceptance criteria for the device was met upon product release, the adhesive component may not meet 3m's performance criteria throughout product shelf life. Additionally, 3m has observed an increase in reported adhesive related skin injuries for these affected lots.
Recall start date: October 7, 2022