3M Surgical Clipper by Remington (2019-01-12)
- Starting date:
- January 12, 2019
- Posting date:
- January 31, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68956
Affected Products
3M Surgical Clipper by Remington
Reason
New caution statement added regarding use in mobile/field emergency medical service (EMS) situations. To reduce the risk of burns associated with overheating, charge the handle using a standard facility 120vac outlet. In EMS situations, use only with a pure sine wave power inverter.
Affected products
3M Surgical Clipper by Remington
Lot or serial number
Not applicable
Model or catalog number
9602
Companies
- Manufacturer
-
3M Health Care
2510 Conway Avenue
St. Paul, Minnesota
55144
UNITED STATES