Issue:
Product Safety, Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-30513

Reason

In the 3D top ceiling supports, screws may become loose when preventative maintenance instructions are not followed. This may cause the rollers to detach from the ceiling track.

 

Affected products

A) 3D Ceiling Stand

B) 3D Ceiling Stand

C) 3D Ceiling Stand

Lot or serial number

A) N/A B) N/A C) N/A

Model or catalog number

A) 3070039

B) 3070013

C) 3070021, 3074486

Companies

Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY