This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
A) 3D Ceiling Stand B) 3D Ceiling Stand C) 3D Ceiling Stand
- Starting date:
- November 23, 2005
- Posting date:
- April 3, 2006
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-30318
Reason
Need to ensure that device is tightened correctly during installation because if not correct, then the rollers may detach from the ceiling rails which may cause the teleskop carriage to loosen and/or fall.
Affected products
A) 3D Ceiling Stand B) 3D Ceiling Stand C) 3D Ceiling Stand
Lot or serial number
A) N/A B) N/A C) N/AModel or catalog number
A) 3070039 B) 3070013 C) 3070021, 3074486Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY